VITAPIA 7000K

Compare
Categories: ,

Description

Vitapia 7000k

  • 12.1″ High brightness TFT LCD display
  • Portable & streamline handle design
  • Big figure display
  • ECG waveforms of 7 -lead display in one screen
  • 72 hours review trend
  • 480 hours review trend with external memory
  • Compatible and self-identify 3/5-lead ECG cable
  • NIBP dual over pressure protection
  • Isolated floating, anti-defibrillation protected and anti-high frequency interference
  • Networking capacity with standard parameter and configuration
  • Optical parameter & configuration : 2-IBP, 2-TEMP, Recorder, 12-lead ECG, EtCO2
  • Central Networl S/W, Wireless Network, VGA, Touch LCD
Technical Specifications

Classfication
Anti-electrochock type Class lequipment and internal powered equipment
EMC type Class A
Anti-electroshock degree ECG(RESP), SPO2, NIBP, IBP, TEMP, CO2, CF
data_01.gif
Specifications
Size Monitor 310 X 140 X 263 mm
Weight Monitor 3.8kg
Temperature Working 5 ~ 40℃
Transport / Storage -40 ~ +55℃
Humidity Working 30% ~ 75%
Transport / Storage ≤ 95% (no coagulate)
Altitude Working 700hPa ~ 1060hPa
Transport / Storage -500hPa ~ 1060hPa
Power Supply 100 ~ 240 VAC, 50/60Hz, Pmax = 150VA, FUSE T1.6A
Display Device 12.1 in. color TRT, 3 LED / 800 X 600 dots
Touch LCD(Option) viewing angle ( up : 70º, down : 70º, Left : 50º, Right : 60º and the best angle is 60º)
Messages
– 8 Wave form Maximum
– 1 Power LED (Green)
– 3 Soung Mode corresponding Alarm Mode
– 1 Alarm LED (Yellow / Red)
– 1 Battery Charge LED (Yellow)
Signal Interface ECG Output BNC
Amplitude 1 mV
Accuracy < 5%
Impedance 50Ω
Signal Delay < 20ms
Battery
– Rechargeble 3.7 A / Hr 7.4 V Li battery
– Operating time under the normal use and full charge greater than 120 min.
– Operating time after the first alarm of low battery will be about 5 min.
ECG Trend Recall
– Short : 1hrs, 1 second Resolution
– Long : 72hrs, 1 min Resolution
Alarm Event Recall 71 alarm events of all parameters and 8/16/32 seconds of corresponding waveform.
NIBP Measurement
Recall
At least 400 NIBP measurement data
SD card – 72 hrs ECG waveform – 480 hrs TREND review
Lead mode 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection Ⅰ, Ⅱ, Ⅲ
Waveform 2CH
Lead mode 3 Leads (R, L, F or RA, LA, LL)
Lead selection
Ⅰ, Ⅱ, Ⅲ
Waveform 1ch
Gain
x 2.5mm / mV, x5.0mm/mV, x10mm/mV, x20mm/mV
HR
– Measuring Range : 15 ~ 300 bpm
– Alarm Range(PED/NEO) : 15~350bpm
– Resolution : 1bpm
– Alarm Rangle(ADU) : 15~300bpm

– Accuracy : ±1% or ±1bpm, which great
– Sensitivity : >200uV P-P
Differental input
Impedance
>5MΩ
CMRR – Monitor : ≥100dB

– Operation : ≥100dB
– Diagnosis : ≥60dB
Electode offset
potential
±300mV
Leakage Current <10uA
Baseline Recovery <5s After Defi
ECG Signal Range ±8mV(Vp-p)
Bandwidth
– Surgery : 1 ~ 20Hz (+0.4dB, -3dB)
– Monitor : 0.5Hz ~ 40Hz (+0.4dB, -3dB)
– Diagnostic : 0.05Hz~75Hz (+0.4dB, -3dB), 76Hz~150Hz(+0.4dB, -4.5dB)
Calibration Signal 1mV(Vp-p), ±5% Accuracy
ST Segment
Monitoring Range
Measure Range : -0.6 ~ +0.8mV
Alarm Range : -2.0 ~ + 2.0mV
ARR Detecting
Type : ASYSTOLE, VFIB/VTC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC,
TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm : Available
Review : Available
RESPARATION Method Impedance between R-F(RA-LL)
Differential Input
Impedance
>2.5MΩ
Measuring
Impedance Range
0.3 ~ 5.0Ω
Base line Impedance
Range
0.1 ~ 2.5KΩ
Bandwidth 0.3 ~ 2.5Hz
Resp. Rate
– Measuring Range : 0 ~ 120rpm
– Alarm Range(PED/NEO) : 0 ~ 150 rpm
– Accuracy : ±2rpm
– Alarm Page(ADU) : 0 ~ 120rpm
– Resolution : 1rpm
Apnea Alarm 10 ~ 40S
NIEP Oscillometric
Mode Manual, Auto, STAT
Measuring
Interval in AUTO
Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480, 960 min
Measuring
Interval in STAT
Mode
5min
Alarm Type SYS, DIA, MEAN
Measuring and alarm Range
Adult mode
– SYS : 40 ~ 270mmHg
– DIA : 10 ~ 215mmHg
– MEAN : 20 ~ 235mmHg
Pediatric mode
– SYS : 40 ~ 200mmHg
– DIA : 10 ~ 150mmHg
– MEAN : 20 ~ 165mmHg
Neonatal mode
– SYS : 40 ~ 135mmHg
– DIA : 10 ~ 100mmHg
– MEAN : 20 ~ 110mmHg
Resolution Pressure : 1mmHg
Accuracy
– Pressure(Maximum Mean error) : ±5mmHg
– Pressure(Maximum Standard deviation) : ±8mmHg
Overpressure
protection
– Adult mode : 297±3mmHg
– Pediatric mode : 240±3mmHg
– Neonatal mode : 147±3mmHg
SpO2
Measuring
Range
0 ~ 100%
Alarm Range 0 ~ 100%
Resolution 1%
Accuracy
– 70 ~ 100% ±2%
– 0 ~ 69% unspecified
Alarm Delay about 1sec
Actualization interval 10sec
Pulse Rate
– Measuring and Range : 0 ~ 250bpm
– Resolution : 1bpm
– Accuracy : ±2bpm
TEMPERATURE Channel 2
Measuring and
Alarm Range
0 ~ 50℃
Resolution 0.1℃
Accuracy ±0.2℃
Actualization
interval
about 1sec
Average Time
Constant
<10sec
IBP (Option) Channel 2
Label ART, PA, CVP, RAP, ICP, P1, P2
Measuring
range
-10 ~ +300mmHg
Alarm range -50 ~ +350mmHg
Press Sensor
– Sensitivity : 5uV/V/mmHg
– Impedance : 300 ~ 3000Ω
Resolution 1mmHg
Accracy ±2% or 1mmHg which great
Actualization
Interval
about 1sec
CO2
(Option)
Method Infra-red Absorption Technique
Measuring
mode
Mainstream and Sidestream
Side-stream
mode sampling
pas flow rate
50ml/Min. ± 10ml/Min.
Measuring range
CO2 : 0 ~ 150 mmHg
INSCO2 : 0 ~ 150 mmHg
AwRR : 2 ~ 150 rpm
Resolution
CO2 : 0.1mmHg (0 ~ 69mmHg), 0.25mmHg (70 ~ 150mmHg)
INSCO2 : 0.1mmHg (0 ~ 69mmHg), 0.25mmHg (70 ~ 150mmHg)
Accuracy
– CO2 : ±2mmHg, 0 ~ 40mmHg
±5% of reading, 41 ~ 70mmHg
±8% of reading, 71 ~ 100mmHg
±10% of reading, 101 ~ 150mmHg
– AqRR : ±1rpm
Pace maker
(Option)
Availalbe
Initialization
Time
– Mainstrea : Capnogram displayed in less than 15sec at an ambient Temp. of 25 ℃,
full sepcification within 2 min
– Sidestrea : Capnogram displayed in less than 20sec at an amblient Temp. of 25℃,
full sepcification within 2 min
– Mainstream Rise Time : Less then 60ms-Adult Reusable or Single-Patient Use
Airway Adaptor
Sidestream
Delay Time
about 1sec
Alarm range
2 ~ 3sec
Suffocation
Alarm Delay
CO2 : 0 ~ 150mmHg